Near-infrared holography extends lensless microscopy into thicker tissue
Researchers at Warsaw University of Technology say near-infrared illumination and a larger sample-sensor gap let lensless holographic microscopy recover detail through tissue-like scattering layers far thicker than visible light can handle. The study, published in Opto-Electronic Advances on Sept. 20, 2026, points to a faster, wider-field way to image thick biological samples without lenses or labels.
Why it matters: - Lensless holographic microscopy has been limited to thin, weakly scattering samples because strong scattering destroys the interference pattern it needs to reconstruct an image. - The new approach extends that method into the multiple-scattering regime, opening a path to wider-field imaging of thick biological tissue with simpler hardware. - The work also shows that ordinary silicon sensors can still support near-infrared holography at very low sensitivity, which could reduce the need for specialized infrared components.
What happened: - Researchers at Warsaw University of Technology tested lensless digital holographic microscopy across six wavelengths from 480 to 1100 nm. - The study used tissue-mimicking milk layers and calibrated targets to measure how much detail remained recoverable as scattering increased. - The results were published online in Opto-Electronic Advances on Sept. 20, 2026. - The paper is titled "Near-infrared lensless holographic microscopy on a standard sensor for label-free mesoscale imaging in multiple scattering." - The DOI is 10.29026/oea.2026.260088.
The details: - At visible wavelengths, the system became fully unresolved once scattering-layer thickness reached 180 to 360 µm, depending on wavelength. - At 1100 nm, resolvable features survived through about 1.4 mm of tissue-like medium. - That depth is nearly four times greater than the thickness reached at 632 nm. - The team used a standard CMOS sensor, even though its quantum efficiency at 1100 nm was only 0.19%. - The lensless setup delivered a field of view of about 116 mm2 without raster scanning or image stitching. - Increasing the sample-sensor distance from about 3 mm to 12 mm improved lateral resolution from roughly 11 µm to 5.5 µm. - The team says the longer propagation distance lets diffuse background spread faster than the useful interference signal, making fringes easier to detect. - At 1100 nm, the system produced label-free amplitude and phase images of uncleared mouse liver sections up to 60 µm thick and brain sections up to 250 µm thick. - The study was led by first author Dr. Emilia Wdowiak, with collaborators from the Nencki Institute of Experimental Biology, the Medical University of Warsaw, the Maria Sklodowska-Curie National Research Institute of Oncology, and the Centre of Postgraduate Medical Education. - The research was supported by Poland's National Centre for Research and Development through the LIDER XIV programme. - Equipment was co-funded by the Warsaw University of Technology Excellence Initiative: Research University program.
Between the lines: - The key advance is not a new microscope body, but a new operating regime for an existing one. - Near-infrared light preserves more usable information in scattering media, and lensless imaging can exploit that with a standard sensor architecture. - The finding that larger sample-sensor gaps can improve resolution runs against a common rule in lensless imaging and suggests geometry should be tuned for scattering conditions, not just minimized. - The work stays in the in vitro and ex vivo lane; it does not claim deep imaging inside living tissue.
What's next: - The team says future work could combine distance-aware acquisition with scattering-aware or physics-informed reconstruction. - More sensitive detectors could further improve performance in low-photon conditions. - The approach could support faster tissue survey, computational imaging through complex media, and compact instruments for labs that need scalable imaging systems.
The bottom line: - Near-infrared lensless holography on a standard sensor can push label-free microscopy much deeper into scattering tissue than visible light, while preserving the wide field and simplicity that make the method attractive.
Disclaimer: This article was produced by AGP Wire with the assistance of artificial intelligence based on original source content and has been refined to improve clarity, structure, and readability. This content is provided on an “as is” basis. While care has been taken in its preparation, it may contain inaccuracies or omissions, and readers should consult the original source and independently verify key information where appropriate. This content is for informational purposes only and does not constitute legal, financial, investment, or other professional advice.
Sign up for:
Poland Daily News
The daily local news briefing you can trust. Every day. Subscribe now.
Check Your Email!
We sent a one-time activation link to: .
Confirm it's you by clicking the email link.
If the email is not in your inbox, check spam or try again.
Welcome back!
is already signed up. Check your inbox for updates.